Is there a quick test reprocessed flexible endoscopes?

Immersion nutrient media such as BODE Dip Slides may be used for this purpose. Application: rinse the respective channels with 20 ml of sterile saline solution and collect the liquid with a sterile Dip Slide tube. Immerse BODE Dip Slide carrier for 2 – 3 min, discard liquid and incubate carrier in the respective tube at 20 – 30 °C for 2 -3 days. Evaluation is carried out on the basis of the scheme. However, this method does not replace the recommendations of the RKI.